Electroanalytical Methods
CSIR NET: potentiometry and ion-selective electrodes, conductometric titration, coulometry, polarography, Ilkovic equation, cyclic voltammetry, amperometry.
- 23 min
On this page · 14 sections
Prerequisites
- Spectroscopic Methods of Analysis: calibration, standard addition and detection limits apply equally to electrodes.
- Acid–Base Concepts: pH and weak-acid equilibria, which shape potentiometric and conductometric titration curves.
- Non-aqueous Solvents: titrations of very weak bases in glacial acetic acid are followed potentiometrically.
- BSc electrochemistry: cell notation, standard potentials, ΔG = −nFE.
The Hook: A Drop of Mercury That Won a Nobel Prize
graph TD
A[Electroanalytical methods] --> B[Zero current
measure potential]
A --> C[Current flows]
A --> G[AC, no net reaction
measure conductance]
B --> B1[Potentiometry
pH, ion-selective electrodes,
potentiometric titrations]
C --> D[Measure charge
whole analyte reacted]
C --> E[Measure current
small fraction reacted]
D --> D1[Coulometry
electrogravimetry]
E --> E1[Voltammetry and polarography
sweep potential]
E --> E2[Amperometry
fixed potential]
G --> G1[Conductometric titrations]1. Electrodes and the Nernst Equation
For a half-reaction \( \text{Ox} + n\text{e}^- \rightleftharpoons \text{Red} \) at 25 °C:
$$\boxed{E = E^\circ - \frac{0.05916}{n}\log\frac{a_{\text{Red}}}{a_{\text{Ox}}}}$$In real solutions the formal potential \( E^{\circ\prime} \) (measured in a stated medium, with concentrations in place of activities) is used. Example: Fe³⁺/Fe²⁺ is +0.771 V standard but about +0.68 V in 1 M H₂SO₄.
Reference electrodes
| Electrode | Half-cell | Potential vs SHE at 25 °C |
|---|---|---|
| Standard hydrogen (SHE) | Pt, H₂ (1 bar) / H⁺ (a = 1) | 0 by definition |
| Saturated calomel (SCE) | Hg / Hg₂Cl₂(s) / KCl (saturated) | +0.241 V |
| Silver–silver chloride | Ag / AgCl(s) / KCl (saturated) | +0.197 V |
Both practical references are electrodes of the second kind: a metal in contact with its sparingly soluble salt and a solution of the common anion, so their potential is fixed by the chloride activity. A potential measured against SCE is converted to SHE by adding 0.241 V.
Indicator electrodes
| Kind | Example | Responds to |
|---|---|---|
| First kind | Ag wire in Ag⁺ solution | activity of its own cation |
| Second kind | Ag/AgCl in Cl⁻ solution | the anion of the sparingly soluble salt |
| Redox (inert) | Pt or Au | ratio Ox/Red of any couple in solution (Fe³⁺/Fe²⁺, Ce⁴⁺/Ce³⁺) |
| Membrane | glass electrode, ion-selective electrodes | a specific ion, through a membrane potential |
2. Potentiometry
The glass electrode
A thin glass membrane separates an internal solution of fixed pH from the test solution. Ion exchange at the hydrated glass surfaces sets up a membrane potential that depends on the outer H⁺ activity:
$$E_{\text{cell}} = K - 0.05916\,\text{pH} \quad (25\ ^\circ\text{C})$$K is unknown and drifts, so the electrode is calibrated with standard buffers before use. Two errors to know:
- Alkaline (sodium) error: at high pH and high Na⁺, the glass responds to Na⁺ as well; the measured pH is lower than the true pH.
- Acid error: in very strongly acidic solutions the measured pH is higher than the true pH.
Ion-selective electrodes (ISEs)
| Type | Example | Membrane | Main interferences |
|---|---|---|---|
| Glass | pH; Na⁺ (modified glass) | silicate glass | Na⁺ at high pH (for pH glass) |
| Crystalline (solid-state) | F⁻ | single crystal of LaF₃ doped with EuF₂ | OH⁻ |
| Crystalline | Cl⁻, Br⁻, I⁻, S²⁻, Ag⁺ | Ag₂S or AgX pressed pellet | ions forming less soluble Ag salts |
| Liquid membrane / polymer | K⁺ | valinomycin (neutral ionophore) in PVC | Rb⁺ and Cs⁺ (similar size), NH₄⁺; very little response to Na⁺ |
| Liquid membrane | Ca²⁺ | calcium dialkyl phosphate ion exchanger | other divalent ions |
| Gas-sensing | CO₂, NH₃ | gas-permeable membrane over a pH electrode in a thin film of bicarbonate or NH₄Cl | volatile acids or bases |
The ideal response is Nernstian: 59.16/z mV per tenfold change in activity at 25 °C (59 mV for H⁺ or K⁺, 29.6 mV for Ca²⁺, −59 mV for F⁻ since it is an anion).
Real electrodes respond partly to other ions. The Nikolsky–Eisenman equation for primary ion i of charge \( z_i \) and interferent j of charge \( z_j \):
$$\boxed{E = \text{const} + \frac{0.05916}{z_i}\log\left(a_i + \sum_j k_{ij}\,a_j^{\,z_i/z_j}\right)}$$The selectivity coefficient \( k_{ij} \) should be much less than 1. If \( k_{\text{Na,K}} = 1.0 \times 10^{-3} \), then 0.10 M K⁺ adds the same response as \( 1.0 \times 10^{-4} \) M Na⁺: a sample with \( a_{\text{Na}} = 1.0 \times 10^{-4} \) M would read as if it contained twice that.
Potentiometric titrations
The cell potential is recorded against titrant volume; the end point is the steepest point of the sigmoid.
- First derivative: plot ΔE/ΔV against mean volume; the maximum is the end point.
- Second derivative: Δ²E/ΔV² changes sign at the end point.
- Gran plot: before the end point a function proportional to the amount of analyte left, for a strong acid \( (V_0 + V)\,10^{-\text{pH}} \) computed from the measured potential, falls linearly with titrant volume V and extrapolates to zero at the end point. It uses points away from the equivalence region, which helps with dilute or weak systems.
For a redox titration in which both couples are reversible and no H⁺ appears in the overall reaction, the potential at equivalence is
$$\boxed{E_{\text{eq}} = \frac{n_1 E_1^{\circ\prime} + n_2 E_2^{\circ\prime}}{n_1 + n_2}}$$For Fe²⁺ titrated with Ce⁴⁺ in 1 M H₂SO₄ (formal potentials about +0.68 V and +1.44 V, both one-electron): \( E_{\text{eq}} = (0.68 + 1.44)/2 = 1.06 \) V. The steep jump is symmetric about \( E_{\text{eq}} \) only when \( n_1 = n_2 \).
Potentiometric end points work in coloured or turbid solutions and in non-aqueous solvents, where indicators fail.
3. Conductometry
Conductance G = 1/R is measured with an AC bridge (to avoid electrolysis) in a cell of constant \( K_{\text{cell}} = l/A \): conductivity \( \kappa = G \times K_{\text{cell}} \), molar conductivity \( \Lambda_m = \kappa/c \).
Strong electrolytes follow the Kohlrausch law \( \Lambda_m = \Lambda_m^\circ - K\sqrt{c} \), and limiting molar conductivities are additive over ions (Kohlrausch’s law of independent migration). Approximate limiting molar ionic conductivities at 25 °C, in S cm² mol⁻¹:
| Ion | λ° | Ion | λ° |
|---|---|---|---|
| H⁺ | ≈ 350 | OH⁻ | ≈ 199 |
| K⁺ | ≈ 73.5 | Cl⁻ | ≈ 76.3 |
| NH₄⁺ | ≈ 73.5 | NO₃⁻ | ≈ 71.4 |
| Ag⁺ | ≈ 61.9 | CH₃COO⁻ | ≈ 40.9 |
| Na⁺ | ≈ 50.1 |
H⁺ and OH⁻ are exceptionally mobile (proton hopping, the Grotthuss mechanism). That one fact predicts almost every titration curve.
Conductometric titration curves
Conductance is plotted against titrant volume (titrant concentrated, so dilution is small, or a correction applied). Each branch is close to a straight line; the end point is the intersection.
| Titration | Before end point | After end point | Shape |
|---|---|---|---|
| Strong acid (HCl) with strong base (NaOH) | falls steeply: fast H⁺ replaced by slow Na⁺ | rises: excess OH⁻ (and Na⁺) | V |
| Weak acid (CH₃COOH) with strong base (NaOH) | small initial dip (common-ion suppression of the little ionisation), then gentle rise as CH₃COO⁻ and Na⁺ form | steeper rise from OH⁻ | two rising lines, break at end point |
| Weak acid with weak base (CH₃COOH with NH₃) | rises as NH₄⁺ and CH₃COO⁻ form | almost flat: excess NH₃ is barely ionised | rise then plateau: sharp end point |
| Mixture HCl + CH₃COOH with NaOH | falls (HCl first), then gentle rise (acetic acid) | steep rise | three branches, two end points |
| Precipitation: KCl with AgNO₃ | nearly flat, slight fall: Cl⁻ replaced by NO₃⁻ (similar λ), Ag⁺ removed as AgCl | rises: excess Ag⁺ and NO₃⁻ | flat then rise |
| Strong acid with weak base (HCl with NH₃) | falls (H⁺ replaced by NH₄⁺) | flat (excess NH₃ un-ionised) | fall then plateau |
Strengths: works for very weak acids, coloured or dilute solutions, and mixtures. Weakness: a large background of unrelated electrolyte swamps the change.
4. Electrogravimetry and Coulometry
Both are exhaustive methods: the whole of the analyte is electrolysed, so the result follows from Faraday’s law with no calibration.
$$\boxed{n_{\text{analyte}} = \frac{Q}{nF}} \qquad Q = I\,t \ (\text{constant current}) \quad \text{or} \quad Q = \int I\,dt$$F = 96 485 C mol⁻¹.
Electrogravimetry
The analyte is deposited on a weighed electrode (Cu on a Pt gauze cathode) and weighed again. Controlling the cathode potential (rather than the cell voltage) lets one metal be deposited while another with a more negative reduction potential stays in solution. A deposit is adherent when deposition is slow and H₂ evolution is avoided; nitrate is sometimes added as a cathodic depolariser that is reduced in preference to H⁺.
Coulometry
- Controlled-potential coulometry: the working electrode is held at a potential where only the analyte reacts; current decays exponentially as the analyte is used up, and the integrated charge gives the amount.
- Constant-current coulometry (coulometric titration): a constant current generates a titrant at the electrode, which reacts with the analyte. Time to the end point × current = charge. The requirement is 100% current efficiency: every electron must go to the intended reaction.
| Electrogenerated reagent | Generated from | Analyte examples |
|---|---|---|
| I₂ | I⁻ at anode | As(III), S₂O₃²⁻, H₂S, water (Karl Fischer) |
| Br₂ | Br⁻ at anode | phenols, anilines, alkenes, As(III) |
| OH⁻ | water at cathode | acids |
| H⁺ | water at anode | bases |
| Ag⁺ | Ag anode | halides, thiols |
| Ce⁴⁺, Fe²⁺ | Ce³⁺, Fe³⁺ | reductants, oxidants |
Coulometric titrations avoid standardising a titrant and can use unstable reagents (Br₂, Cl₂, Ti³⁺) made on demand. Coulometric Karl Fischer titration is the standard method for trace water.
5. Voltammetry and Polarography
The set-up
A three-electrode cell: a small working electrode where the reaction of interest occurs, a reference electrode (SCE or Ag/AgCl) against which its potential is controlled, and an auxiliary (counter) electrode that carries the current, so that no current flows through the reference. Only a tiny fraction of the analyte reacts: voltammetry is non-exhaustive.
In polarography the working electrode is the dropping mercury electrode (DME):
- a fresh, reproducible surface with every drop, so no poisoning or memory;
- a high overpotential for H₂ evolution on mercury, so reductions can be studied at quite negative potentials (even alkali metals in a tetraalkylammonium supporting electrolyte);
- but a limited anodic range, because mercury itself oxidises at modestly positive potentials, and mercury is toxic.
The currents in a polarogram
| Current | Cause | How it is handled |
|---|---|---|
| Residual | charging of the electrical double layer of each new drop (capacitive) plus traces of reducible impurities | subtracted; pulse methods reduce it |
| Migration | electrostatic attraction of the charged analyte to the electrode | eliminated by a large excess (about 50–100 times) of inert supporting electrolyte (KCl, tetraalkylammonium salts), which carries almost all the migration current |
| Diffusion | analyte reaching the electrode by diffusion through the depletion layer | the analytical signal |
| Limiting | plateau where every analyte molecule arriving is reduced | limiting − residual = diffusion current \( i_d \) |
| Maxima | streaming of solution round the drop gives a hump above the plateau | suppressed with a little gelatin or Triton X-100 |
Dissolved oxygen gives two waves of its own (O₂ → H₂O₂, then H₂O₂ → H₂O) and must be removed by bubbling N₂ or Ar before the scan.
The Ilkovic equation
$$\boxed{i_d = 708\,n\,D^{1/2}\,m^{2/3}\,t^{1/6}\,C} \quad (\text{maximum current at the end of drop life})$$with \( i_d \) in μA, D in cm² s⁻¹, m the mercury flow rate in mg s⁻¹, t the drop time in s and C in mmol L⁻¹. For the current averaged over the drop life the constant is 607. Consequences:
- \( i_d \propto C \): the basis of quantitative polarography.
- \( i_d \propto D^{1/2} \) and \( \propto n \): a two-electron reduction gives about twice the wave height of a one-electron reduction at the same concentration.
- The product \( m^{2/3} t^{1/6} \) depends on the height h of the mercury column: m ∝ h and t ∝ 1/h, so \( m^{2/3}t^{1/6} \propto h^{2/3}h^{-1/6} = h^{1/2} \). A diffusion current is proportional to √h. A kinetic (reaction-controlled) current is independent of h; an adsorption current is proportional to h. Varying h diagnoses the type of current.
The wave equation and the half-wave potential
For a reversible reduction where both Ox and Red are soluble (Red dissolved in the mercury or solution):
$$\boxed{E = E_{1/2} + \frac{0.05916}{n}\log\frac{i_d - i}{i}}$$At \( i = i_d/2 \), \( E = E_{1/2} \), the half-wave potential. It is independent of concentration and characteristic of the analyte in a given medium, so it identifies the species, while \( i_d \) measures how much. A plot of E against \( \log[(i_d - i)/i] \) is a straight line of slope 0.05916/n V if the reduction is reversible: 59 mV for n = 1, 30 mV for n = 2. A larger slope signals an irreversible electrode process.
Complexation shifts the wave (Lingane)
If the metal ion is complexed, \( \text{ML}_p \) is harder to reduce than the aqua ion and the wave moves to more negative potential:
$$\boxed{(E_{1/2})_{c} - (E_{1/2})_{s} = -\frac{0.05916}{n}\log K_f - p\,\frac{0.05916}{n}\log[\text{L}]}$$(assuming similar diffusion coefficients for free and complexed ion). Plotting \( E_{1/2} \) against log[L] gives a line of slope \( -p(0.05916/n) \): the coordination number p. The intercept at log[L] = 0 gives the formation constant \( K_f \).
Modern variants
| Technique | Idea | Advantage |
|---|---|---|
| Normal / differential pulse polarography | potential pulses applied late in the drop life; current sampled after the charging current has decayed | detection limits about 100 to 1000 times lower than DC polarography; differential pulse gives a peak at about \( E_{1/2} \) |
| Anodic stripping voltammetry (ASV) | metal ions (Pb²⁺, Cd²⁺, Cu²⁺, Zn²⁺) pre-concentrated by reduction into a hanging mercury drop or mercury film for minutes with stirring, then stripped (re-oxidised) by an anodic scan | pre-concentration gives the lowest detection limits of any common electrochemical method; several metals in one scan |
| Cathodic stripping voltammetry | analyte deposited as an insoluble mercury or silver salt by oxidation, then reduced | halides, sulfide |
| Cyclic voltammetry | triangular potential sweep at a stationary electrode | mechanism and reversibility rather than routine quantitation |
Cyclic voltammetry
The potential is swept linearly from an initial value past \( E^{\circ\prime} \) and back. A reduction peak (cathodic, \( i_{pc} \) at \( E_{pc} \)) appears on the forward scan and, if the product is stable, an oxidation peak (\( i_{pa} \) at \( E_{pa} \)) on the return.
For a reversible (fast electron transfer, Nernstian) couple at 25 °C:
| Criterion | Value |
|---|---|
| Peak separation \( \Delta E_p = E_{pa} - E_{pc} \) | about 59/n mV (more precisely 2.22RT/nF ≈ 57/n mV) and independent of scan rate |
| Peak current ratio \( i_{pa}/i_{pc} \) | 1 |
| \( E^{\circ\prime} \) | \( \approx (E_{pa} + E_{pc})/2 \) |
| Peak current | \( \propto v^{1/2} \), given by the Randles–Sevcik equation |
with \( i_p \) in A, A in cm², D in cm² s⁻¹, C in mol cm⁻³ and v in V s⁻¹.
Diagnoses:
- \( \Delta E_p \) growing with scan rate: quasi-reversible (slow electron transfer).
- No return peak at all: irreversible electron transfer or a fast following chemical reaction.
- Return peak smaller than forward (\( i_{pa}/i_{pc} < 1 \)) and recovering at high scan rates: an EC mechanism, where the product decays chemically before it can be re-oxidised; at fast scans there is no time for the chemical step.
- \( i_p \propto v \) rather than \( v^{1/2} \): adsorbed (surface-confined) species.
Amperometric titrations
The working electrode is held at a fixed potential on the limiting-current plateau of a species, and current is recorded against titrant volume. Branches are straight lines.
| Which species is reducible at the chosen potential | Shape |
|---|---|
| Analyte only (Pb²⁺ titrated with SO₄²⁻) | current falls to the end point, then stays at the residual level (L shape) |
| Titrant only (SO₄²⁻-type analyte titrated with Pb²⁺) | residual current until the end point, then rises (reversed L) |
| Both analyte and titrant (Pb²⁺ titrated with dichromate at a potential where both reduce) | falls, then rises (V shape) |
In biamperometric (dead-stop) titrations two identical polarised Pt electrodes with a small applied voltage pass current only when a reversible couple is present in both forms. In the Karl Fischer titration of water, current stays near zero until the first excess of I₂ makes the I₂/I⁻ couple available, then jumps: the end point “stops dead” in reverse.
Memory Tricks
Don’t Confuse
| This | vs | That |
|---|---|---|
| Potentiometry: zero current, measures E | voltammetry: current flows, measures i versus E | |
| Coulometry: exhaustive, absolute (Faraday) | polarography: non-exhaustive, needs calibration | |
| Half-wave potential \( E_{1/2} \): identity | diffusion current \( i_d \): concentration | |
| Diffusion current ∝ √h | kinetic current independent of h; adsorption current ∝ h | |
| Migration current (removed by supporting electrolyte) | diffusion current (the analytical signal) | |
| Alkaline error: pH reads low | acid error: pH reads high | |
| Coulometric titration: reagent generated in situ, counted in coulombs | classical titration: standardised titrant from a burette | |
| CV reversible: ΔEp ≈ 59/n mV, independent of v | quasi-reversible: ΔEp grows with v | |
| Amperometric titration: one polarised electrode, fixed E | biamperometric: two polarised electrodes, small fixed ΔV |
Common Mistakes
Practice
All questions below are original practice questions written for this page. They are not from any CSIR NET paper.
Level 1: Foundation
- What is the potential of the saturated calomel electrode against SHE at 25 °C?
- Why is a large excess of supporting electrolyte added in polarography?
- In cyclic voltammetry of a reversible one-electron couple, what is the expected peak separation?
Answers
- +0.241 V.
- To carry the migration current, so that the analyte reaches the electrode only by diffusion and the limiting current obeys the Ilkovic equation. It also lowers solution resistance.
- About 59 mV (more exactly about 57 mV) at 25 °C, independent of scan rate.
Level 2: Part B Style
Practice Q1. Acetic acid is titrated conductometrically with aqueous ammonia. The plot of conductance against volume of ammonia is best described as:
(a) a V shape (b) a steady rise to the end point, then nearly constant (c) a steep fall, then a steep rise (d) nearly constant, then a steep rise
Answer: (b)
Ammonium acetate, a strong electrolyte, forms from two weak electrolytes, so conductance climbs. After the end point excess NH₃ is a weak base that barely ionises in the presence of NH₄⁺, so conductance stays almost constant. (a) is strong acid–strong base; (d) describes KCl titrated with AgNO₃.
Practice Q2. As(III) is titrated with electrogenerated iodine at a constant current of 20.0 mA. The end point is reached after 482.5 s. The amount of arsenic is (F = 96 485 C mol⁻¹, As = 74.92 g mol⁻¹):
(a) 7.49 mg (b) 3.75 mg (c) 1.87 mg (d) 0.375 mg
Answer: (b) 3.75 mg
Q = 0.0200 × 482.5 = 9.65 C = 1.000 × 10⁻⁴ mol e⁻. Two electrons per I₂, one I₂ per As(III): 5.00 × 10⁻⁵ mol As = 3.75 mg. Option (a) assumes one electron per As.
Practice Q3. In DC polarography, doubling the height of the mercury column multiplies a diffusion-controlled limiting current by about:
(a) 2 (b) 1.41 (c) 1 (d) 0.71
Answer: (b) 1.41
\( m^{2/3}t^{1/6} \propto h^{1/2} \), so the current scales as \( \sqrt{2} = 1.41 \). A kinetic current would be unchanged (c); an adsorption current would double (a).
Practice Q4. A fluoride ion-selective electrode is used with TISAB at pH about 5.5. Which reason for the pH choice is correct?
(a) Below this pH, F⁻ is converted to HF; above it, OH⁻ interferes with the LaF₃ membrane. (b) LaF₃ dissolves above pH 4. (c) The electrode responds to H⁺ below pH 7. (d) Fluoride is oxidised at high pH.
Answer: (a)
HF (pKa about 3.2) is not sensed by the electrode, so acid solutions read low; OH⁻ is the main interferent of LaF₃ at high pH. The window around pH 5 to 5.5 avoids both.
Level 3: Part C Style
Practice Q5. Polarographic half-wave potentials of a metal ion M²⁺ (two-electron, reversible reduction) were measured in solutions of a ligand L. Plotting \( E_{1/2} \) (V vs SCE) against log[L] gives a straight line of slope −0.089 V; extrapolated to log[L] = 0, \( E_{1/2} \) is 0.355 V more negative than for the uncomplexed ion. Which statements are correct?
- The complex is ML₃.
- log K_f of the complex is about 12.0.
- Raising [L] tenfold shifts \( E_{1/2} \) by 0.089 V to more negative potential.
- The complex is reduced more easily than the aqua ion.
(a) 1, 2 and 3 only (b) 1 and 4 only (c) 2 and 3 only (d) 1, 2, 3 and 4
Answer: (a) 1, 2 and 3 only
Lingane: slope = \( -p(0.05916/n) \). With n = 2: \( p = 0.089/0.02958 = 3.0 \): ML₃, statement 1 true.
At log[L] = 0, \( \Delta E_{1/2} = -(0.05916/2)\log K_f = -0.355 \) V, so \( \log K_f = 0.355/0.02958 = 12.0 \): statement 2 true.
A tenfold change in [L] is one log unit, so \( E_{1/2} \) moves by the slope, −0.089 V: statement 3 true.
The wave moves to more negative potential, so the complex is harder to reduce: statement 4 false.
Practice Q6. A cyclic voltammogram of a 1.0 mM solution of a one-electron redox couple is recorded at a disc electrode of area 0.0707 cm² at 0.100 V s⁻¹. Assume D = 7.0 × 10⁻⁶ cm² s⁻¹ and reversible behaviour. Which statements are correct?
- The cathodic peak current is about 16 μA.
- At 0.400 V s⁻¹ the peak current would be about 32 μA.
- At 0.400 V s⁻¹ the peak separation would still be about 59 mV.
- If the ratio \( i_{pa}/i_{pc} \) were 0.5 at 0.100 V s⁻¹ but rose towards 1 at fast scans, a following chemical reaction would be indicated.
(a) 1 and 2 only (b) 1, 2, 3 and 4 (c) 3 and 4 only (d) 1, 3 and 4 only
Answer: (b) 1, 2, 3 and 4
Randles–Sevcik with C = 1.0 × 10⁻⁶ mol cm⁻³:
\( i_p = 2.69 \times 10^5 \times 1 \times 0.0707 \times (7.0 \times 10^{-6})^{1/2} \times 1.0 \times 10^{-6} \times (0.100)^{1/2} = 1.59 \times 10^{-5} \) A ≈ 16 μA: true.
\( i_p \propto v^{1/2} \): four times the scan rate doubles the current, about 32 μA: true.
For a reversible couple ΔEp does not depend on scan rate: true.
A small return peak that recovers at high scan rates is the signature of an EC mechanism (the chemical step has less time to remove the product): true.
Practice Q7. Data near the end point of a potentiometric titration:
| V (mL) | 24.0 | 24.5 | 24.8 | 25.0 | 25.2 | 25.5 | 26.0 |
|---|---|---|---|---|---|---|---|
| E (V) | 0.402 | 0.418 | 0.440 | 0.500 | 0.560 | 0.582 | 0.596 |
Which statements are correct?
- The largest ΔE/ΔV, 0.30 V mL⁻¹, occurs in the intervals centred at 24.9 and 25.1 mL.
- The end point is 25.0 mL.
- The second derivative changes sign between 24.9 and 25.1 mL.
- The end point is 24.8 mL because the largest single jump starts there.
(a) 1, 2 and 3 only (b) 1 and 4 only (c) 2 only (d) 1, 2, 3 and 4
Answer: (a) 1, 2 and 3 only
ΔE/ΔV for successive intervals (mean volume: value): 24.25: 0.032; 24.65: 0.073; 24.9: 0.300; 25.1: 0.300; 25.35: 0.073; 25.75: 0.028 V mL⁻¹.
The first derivative peaks symmetrically about 25.0 mL, and the second derivative is positive before and negative after it. End point 25.0 mL. Statement 4 picks the start of an interval, not the inflection point.
Quick Revision Table
| Idea | Formula or fact |
|---|---|
| Nernst (25 °C) | E = E° − (0.05916/n) log(a_Red/a_Ox) |
| References | SCE +0.241 V; Ag/AgCl (sat. KCl) +0.197 V vs SHE |
| Glass electrode | E = K − 0.05916 pH; alkaline error reads low, acid error reads high |
| ISE | 59.16/z mV per decade; Nikolsky: log(a_i + Σk_ij a_j^(z_i/z_j)) |
| F⁻ ISE | LaF₃ (EuF₂-doped); TISAB at pH ≈ 5–5.5; OH⁻ interferes |
| Redox equivalence | E_eq = (n₁E₁ + n₂E₂)/(n₁ + n₂); Fe²⁺/Ce⁴⁺ ≈ 1.06 V |
| Conductance | λ°(H⁺) ≈ 350, λ°(OH⁻) ≈ 199 S cm² mol⁻¹ dominate |
| Conductometric shapes | SA–SB V; WA–SB rise-steeper rise; WA–WB rise-flat; KCl–AgNO₃ flat-rise |
| Faraday | n = Q/(nF), Q = It; F = 96 485 C mol⁻¹ |
| Coulometric titration | electrogenerated I₂, Br₂, OH⁻, H⁺, Ag⁺; needs 100% current efficiency |
| Ilkovic | i_d = 708 n D^½ m^⅔ t^⅙ C (max); 607 for average; i_d ∝ √h |
| Wave equation | E = E₁/₂ + (0.05916/n) log[(i_d − i)/i] |
| Lingane | ΔE₁/₂ = −(0.05916/n) log K_f − p(0.05916/n) log[L] |
| CV reversible | ΔEp ≈ 59/n mV; i_pa/i_pc = 1; i_p ∝ √v |
| Randles–Sevcik | i_p = 2.69 × 10⁵ n^(3/2) A D^½ C v^½ |
| Stripping | pre-concentrate, then strip; lowest detection limits |
| Amperometric shapes | analyte reducible: L; titrant reducible: reversed L; both: V |
Key Takeaways
What’s Next
- Thermoanalytical Methods: TGA, DTA, DSC and thermometric titrations.
- Back to Spectroscopic Methods of Analysis for photometric titrations, the optical counterpart of the curves above.
- Titrations in glacial acetic acid and liquid ammonia: Non-aqueous Solvents.
- Chapter summary: Analytical Chemistry: Quick Revision.