Analytical Chemistry Formula Sheet and Short Notes

One-page CSIR NET revision: extraction, chromatography, Beer–Lambert, AAS and emission, potentiometry, conductometry, polarography, CV, TGA, DTA and DSC.

  • 7 min read
  • Updated
On this page · 5 sections
  1. Separation methods
  2. Spectroscopic methods
  3. Electroanalytical methods
  4. Thermoanalytical methods
  5. Where to go next

Every formula, curve shape and trap from Separation Methods, Spectroscopic Methods, Electroanalytical Methods and Thermoanalytical Methods, on one page.

Separation methods

Solvent extraction

$$\boxed{q_n = \left(\frac{V_{aq}}{V_{aq} + D V_{org}}\right)^n} \qquad \%E = \frac{100D}{D + V_{aq}/V_{org}}$$
  • K_D: one species, constant. D: all forms, depends on pH and ligands.
  • D = 5, 100 mL water: one 60 mL portion extracts 75%; three 20 mL portions 87.5%.
  • Weak acid: \( D = K_D[\text{H}^+]/([\text{H}^+] + K_a) \); D = K_D/2 at pH = pKa.
  • Chelate: \( \log D = \log K_{ex} + n\log[\text{HL}]_{org} + n\,\text{pH} \). Slope = charge n; pH₁/₂ falls 1 unit per tenfold [HL].
  • Ion association: FeCl₄⁻ into ether from HCl; TBP extracts UO₂(NO₃)₂(TBP)₂; synergism with neutral donors.

Chromatography

QuantityFormula
Retention factork = (t_R − t_M)/t_M
Selectivityα = k_B/k_A (≥ 1)
PlatesN = 16(t_R/W)² = 5.54(t_R/W₁/₂)²
Plate heightH = L/N
van DeemterH = A + B/u + Cu; u_opt = √(B/C); H_min = A + 2√(BC)
ResolutionR_s = 2Δt_R/(W_A + W_B); 1.5 = baseline
PurnellR_s = (√N/4)[(α − 1)/α][k_B/(1 + k_B)]
Kovats\( I = 100\left[n + \dfrac{\log t'_x - \log t'_n}{\log t'_{n+1} - \log t'_n}\right] \); n-alkane = 100 × C
  • A: eddy diffusion (absent in open tubular columns). B/u: longitudinal diffusion (large in GC, tiny in HPLC). Cu: mass transfer.
  • R_s ∝ √L: doubling R_s needs 4 × the column. α is the strongest lever.
  • General elution problem: temperature programming (GC), gradient elution (HPLC).
ModeElutes first
Normal phase (silica)least polar
Reversed phase (C18)most polar; more organic solvent = shorter retention
Size exclusionlargest
Cation exchange with complexing eluent (lanthanides)Lu (La last)
Anion exchange from conc. HClNi (no anionic chloro-complex), then Co, Fe; Zn held longest
Capillary zone electrophoresis (normal EOF)cations, then neutrals together, then anions
GC detectorSeesBlind to
TCDeverything (universal)— (low sensitivity)
FIDorganics (C–H)H₂O, CO₂, CO, N₂, O₂
ECDhalogenated, nitro compoundshydrocarbons
FPD / NPDS, P / N, Pmost others
  • Ion-exchange selectivity: higher charge wins; same charge, smaller hydrated radius wins: Cs⁺ > Rb⁺ > K⁺ > Na⁺ > Li⁺; Ba²⁺ > Sr²⁺ > Ca²⁺ > Mg²⁺; I⁻ > Br⁻ > Cl⁻ > F⁻.

Spectroscopic methods

$$\boxed{A = -\log T = \varepsilon b c} \qquad A_\lambda = b\sum \varepsilon_{i,\lambda} c_i$$
  • T 10% ↔ A 1; T 1% ↔ A 2. Best precision at A = 0.434 (T = 1/e); work at 0.2–0.8.
  • Deviations: real (c above about 0.01 M), chemical (dimerisation, dissociation: Cr₂O₇²⁻/HCrO₄⁻), instrumental (polychromatic light, stray light → negative deviation at high A).
  • Stray light: A’ = −log[(T + s)/(1 + s)]; true 2.00 with 1% stray reads 1.70.
  • Isosbestic point: only two interconverting absorbers.
  • LOD = 3s_blank/slope; LOQ = 10s.
  • Standard addition corrects matrix effects; internal standard corrects fluctuations.
  • Job’s maximum at x_L = n/(n + 1): 0.5 ML, 0.67 ML₂, 0.75 ML₃.
  • Photometric titration shapes: titrant absorbs → flat then rise; product absorbs → rise then flat; analyte absorbs → fall then flat; analyte and titrant → V.

Atomic methods

InterferenceCure
Spectral, backgroundD₂ lamp, Zeeman, Smith–Hieftje correction
Chemical (Ca + phosphate)releasing agent La³⁺/Sr²⁺; protective agent EDTA; N₂O–C₂H₂ flame
Ionisation (alkali metals)suppressor: excess K or Cs
Matrix (viscosity)standard addition, matrix matching
  • AAS: hollow-cathode lamp of the analyte (lines too narrow for a continuum); graphite furnace gives lower LOD; cold vapour for Hg; hydride generation for As, Se, Sb, Bi, Sn.
$$\boxed{\frac{N_j}{N_0} = \frac{g_j}{g_0}e^{-\Delta E/kT}}$$
  • Na 589 nm (ΔE 2.10 eV, g ratio 3): 1.7 × 10⁻⁴ at 2500 K, 2.5 × 10⁻⁴ at 2600 K. Emission is temperature-sensitive; absorption is not.
  • Flame photometry: Na 589, K 766, Li 671 nm. ICP: Ar plasma, multi-element, few chemical interferences, many spectral lines.

Fluorescence

  • F ≈ 2.303kφP₀εbc for A < 0.05; inner-filter bends the plot at high c. More sensitive than absorption.
  • Quenchers: heavy atoms, O₂, paramagnetic species.
$$\boxed{\frac{F_0}{F} = 1 + K_{SV}[\text{Q}] = 1 + k_q\tau_0[\text{Q}]}$$
  • Dynamic: τ falls with [Q], more quenching on heating. Static: τ unchanged, less quenching on heating. k_q above about 10¹⁰ L mol⁻¹ s⁻¹ → static contribution.

Electroanalytical methods

Referencevs SHE (25 °C)
SCE (sat. KCl)+0.241 V
Ag/AgCl (sat. KCl)+0.197 V
  • Nernst: E = E° − (0.05916/n) log(a_Red/a_Ox). ISE slope 59.16/z mV per decade.
  • Glass electrode: alkaline (Na⁺) error reads low; acid error reads high; calibrate with buffers.
  • Nikolsky: E = const + (0.05916/z_i) log(a_i + Σ k_ij a_j^(z_i/z_j)).
  • F⁻ ISE: LaF₃ (EuF₂-doped); TISAB, pH ≈ 5–5.5 (HF below, OH⁻ above); CDTA/citrate frees F⁻ from Al³⁺, Fe³⁺.
  • K⁺: valinomycin. Gas-sensing CO₂, NH₃: pH electrode behind a gas-permeable membrane.
  • Redox equivalence: E_eq = (n₁E₁ + n₂E₂)/(n₁ + n₂); Fe²⁺/Ce⁴⁺ in 1 M H₂SO₄ ≈ 1.06 V.
  • End point: max of ΔE/ΔV; sign change of Δ²E/ΔV²; Gran plot.

Conductometric curves

TitrationShape
HCl with NaOHsteep fall, then rise (V)
CH₃COOH with NaOHslight dip, gentle rise, then steeper rise
CH₃COOH with NH₃rise, then flat
HCl + CH₃COOH with NaOHfall, gentle rise, steep rise
KCl with AgNO₃nearly flat, then rise
HCl with NH₃fall, then flat

λ° (S cm² mol⁻¹): H⁺ ≈ 350, OH⁻ ≈ 199, K⁺ 73.5, Cl⁻ 76.3, NO₃⁻ 71.4, Ag⁺ 61.9, Na⁺ 50.1, CH₃COO⁻ 40.9.

Coulometry and voltammetry

$$\boxed{n_{\text{analyte}} = \frac{Q}{nF}} \qquad F = 96\,485\ \text{C mol}^{-1}$$
  • 20.0 mA × 482.5 s = 9.65 C = 1.00 × 10⁻⁴ mol e⁻ = 5.00 × 10⁻⁵ mol As(III) with electrogenerated I₂.
  • Coulometric titrants: I₂, Br₂, OH⁻, H⁺, Ag⁺, Ce⁴⁺; need 100% current efficiency.
$$\boxed{i_d = 708\,n D^{1/2} m^{2/3} t^{1/6} C} \ (\text{607 for average}) \qquad E = E_{1/2} + \frac{0.05916}{n}\log\frac{i_d - i}{i}$$
  • Units: μA, cm² s⁻¹, mg s⁻¹, s, mM. i_d ∝ √h (diffusion); kinetic current independent of h; adsorption current ∝ h.
  • Supporting electrolyte removes migration current; N₂ removes O₂ (two waves); gelatin/Triton X-100 suppresses maxima.
  • DME: renewable surface, high H₂ overpotential, limited anodic range.
  • Reversible wave: log-plot slope 59/n mV.
  • Lingane: ΔE₁/₂ = −(0.05916/n) log K_f − p(0.05916/n) log[L]: slope → p, intercept → K_f.
  • Pulse methods: lower LOD; DPP gives a peak. ASV: pre-concentrate Pb, Cd, Cu, Zn into Hg, then strip.
CV criterion (reversible, 25 °C)Value
ΔEp≈ 59/n mV (2.22RT/nF ≈ 57/n), independent of v
i_pa/i_pc1
i_p2.69 × 10⁵ n^(3/2) A D^½ C v^½ (A, cm², cm² s⁻¹, mol cm⁻³, V s⁻¹)
  • ΔEp grows with v → quasi-reversible. No return peak → irreversible or fast chemical step. Ratio < 1 recovering at fast scans → EC. i_p ∝ v → adsorbed species.
  • Amperometric titrations: analyte reducible → L; titrant reducible → reversed L; both → V. Karl Fischer: biamperometric (dead-stop).

Thermoanalytical methods

TechniqueMeasuresSees melting?
TGAmassno
DTGdm/dTno
DTAΔT (sample − reference)yes, semi-quantitative
DSCheat flowyes, ΔH from area
CaC₂O₄·H₂O (146.11)Loss (% of original)Residue
−H₂O12.387.7% CaC₂O₄
−CO19.268.5% CaCO₃
−CO₂30.138.4% CaO
  • CO step: endothermic in N₂, exothermic in air (CO burns).
  • CuSO₄·5H₂O: 2 + 2 + 1 H₂O (7.22% each; last one held most firmly); CuSO₄ 63.9%; CuO 31.9%.
  • Faster heating, bigger sample, product gas in purge → steps at higher T. Buoyancy → apparent mass gain.
  • Ca/Mg oxalate mixture: mass loss between CaCO₃ + MgO and CaO + MgO plateaus is CO₂ from CaCO₃ alone.
  • Endothermic: melting, vaporisation, dehydration, most decompositions. Exothermic: crystallisation, oxidation, curing, energetic decompositions.
  • DSC: area = ΔH; baseline ∝ C_p; T_g = step; indium (mp 156.6 °C) calibrates.
$$\boxed{X_c = \frac{\Delta H_m - \Delta H_{cc}}{\Delta H_m^{\circ}}}$$
  • Thermometric titration: temperature vs volume; responds to ΔH, so weak acids such as boric acid give breaks.

Where to go next

Back to the chapter overview: Analytical Chemistry. Acid–base background for extraction and titrations is in Acids, Bases and Non-aqueous Solvents: Quick Revision, and the biochemist’s view of the same separations in Methods in Biology. Subject overview: Chemical Sciences.